MED75y Series Instruments
The MED75y Series Instruments are designed for the high speed testing of medical products ranging in volume from several cubic-centimeters to several liters. InterTech’s patented mass flow test technology results in a cost effective solution to the requirements of high speed testing.
MED75Y-Series Mass Flow Instrumentation Principles
The mass flow transducer is connected to both the reservoir, and the test part. When this circuit is filled with air and isolated from the supply, any leak results in a flow of air from the reservoir to the test part through the transducer. The transducer output provides a direct measurement of leakage.
>MED75Y-Series Technical Specifications & Performance Metrics
- Leak range: 0 - 20 sccm.
- Pressure range: 0 - 100 psig.
- Transducer repeatability: 0.05% F.S.D.
- Transducer sensitivity: 0.05% F.S.D.
- AD Conversion: 24 bit, 400 samples/sec.
- Timer increments: 0.01 sec.
- NIST traceable calibration to less than 5% R&R.
FAQs
MED75y Series Leak Detection Instruments are optimized for the small volumes and high throughput requirements common in medical device manufacturing. Their patented mass flow technology enables highly accurate leak testing while supporting rapid production cycles.
Mass flow technology directly measures air movement through a leak path, providing highly sensitive and repeatable measurements. MED75y Series Leak Detection Instruments use this approach to improve testing accuracy for medical products with strict quality requirements.What medical products commonly use MED75y Series Instruments for leak testing?
MED75y Series Instruments are commonly used to test infusion devices, catheters, fluid delivery systems, diagnostic products, surgical instruments, and other medical components that require precise leak detection and flow verification.
High-speed leak detection systems reduce test cycle times while maintaining accuracy, allowing manufacturers to increase throughput without sacrificing quality. MED75y Series Leak Detection Instruments are specifically designed to support these demanding production environments.
Manufacturers should evaluate part volume, required leak sensitivity, cycle time requirements, regulatory standards, and data collection capabilities when selecting a medical leak testing instrument for production environments.
Consult Our Experts
Please contact InterTech for further information or help with your test application. We are happy to answer any questions you may have. Give us a call at 847-679-3377 or fill out the form and we'll be in touch as soon as possible.
*denotes a required field.